Home Infectious Disease Test Fluorescence Immuno Assay Diagnostic Kit For Cardiac Troponin I
Fluorescence Immuno Assay Diagnostic Kit For Cardiac Troponin I
Fluorescence Immuno Assay Diagnostic Kit For Cardiac Troponin I
Fluorescence Immuno Assay Diagnostic Kit For Cardiac Troponin I
Fluorescence Immuno Assay Diagnostic Kit For Cardiac Troponin I
Fluorescence Immuno Assay Diagnostic Kit For Cardiac Troponin I
Fluorescence Immuno Assay Diagnostic Kit For Cardiac Troponin I
Fluorescence Immuno Assay Diagnostic Kit For Cardiac Troponin I
Fluorescence Immuno Assay Diagnostic Kit For Cardiac Troponin I

Fluorescence Immuno Assay Diagnostic Kit For Cardiac Troponin I

We supply Fluorescence Immuno Assay Diagnostic Kit For Cardiac Troponin I,microbiology infectious diseases,national institute of allergy,infectious diseases.
Product Details

Diagnostic Kit for Cardiac Troponin l

(Whole blood/Serum/Plasma)( Fluorescence Immunochromatographic Assay)

PRINCIPLE

The strip has anti-FOB coating antibody on test region, which is fastened to membrane chromatography in advance. Lable pad is coated by fluorescence labeled anti-FOB antibody in advance. When testing positive sample, the FOB in sample can be mixed with fluorescence labeled anti-FOB antibody, and form immune mixture. As the mixture is allowed to migrate along the test strip, the FOB conjugate complex is captured by anti-FOB coating antibody on the membrane and forms complex. The fluorescence intensity is positively correlated with the FOB content. The FOB in sample can be detected by fluorescence immunoassay analyzer.

Test Procedure:

1.Lay aside all reagents and samples to room temperature.

2.Open the Portable Immune Analyzer(WIZ-A101), enter the account password login according to the operation method of the instrument, and enter the detection interface.

3.Scan the dentification code to confirm the test item.

4.Take out the test card from the foil bag.

5.Insert the test card into the card slot, scan the QR code, and determine the test item.

6.Remove the cap from the sample tube and discard the first two drops diluted sample, add 3 drops (about 100uL) no bubble diluted sample verticaly and slowly into sample well of the card with provided dispette.

7.Click the "standard test" button, after 15 minutes, the instrument will automatically detect the test card, it can read the results from the display screen of the instrument, and record/print the test results.

8.Refer to the instruction of Portable Immune Analyzer(WIZ-A101).

QQ截图20220323134018.jpg

Clinical significance:

1.Auxiliary diagnosis of myocardial infarction.

2.The preferred marker for detection of myocardial injury.

3.ldentify the reason for chest pain.

4.Diagnosis and risk stratification of acute coronary syndrome.

5.The best marker was marker was ACS surgey.

Features:

Quick: whole blood,serum or plasma detection,results come out in 15 mins.

Accuracy : good correlation with the results of chemiluminescence detection(R^2>0.95).

Application: simple operation,suitable for bedside,outpatient.

ICU,outdoor and so on.


Model NumberCTNIPacking25 Tests/ kit,   20kits/CTN
NameDiagnostic Kit   for Cardiac Troponin I ( Fluorescence Immuno Assay)Instrument    classificationClass II
FeaturesHigh   sensitivity, Easy opeationCertificateCE/ ISO13485
SpecimenSerum, PlasmaShelf lifeTwo Years
Accuracy> 99%TechnologyQuantitative   kit
Storage2′C-30′CTypePathological   Analysis Equipments